GSO IEC 61223-3-7:2024 Scope

IEC 61223-3-7:2021 applies to DENTAL CONE-BEAM COMPUTED TOMOGRAPHY X-RAY EQUIPMENT, hereafter also called DENTAL CBCT EQUIPMENT, that conforms to IEC 60601-2-63:2012 and IEC 60601-2-63:2012/AMD1:2017.
This document applies to ACCEPTANCE TESTS and CONSTANCY TESTS on DENTAL CONE-BEAM COMPUTED TOMOGRAPHY X-RAY EQUIPMENT.
The aim of ACCEPTANCE TESTS is to verify compliance of the installation or MAJOR SERVICE ACTION with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning.
The requirements specified in this document are minimal requirements. The MANUFACTURER can establish criteria for the tests described here that exceed the levels contained in this document.
CONSTANCY TESTS are performed to ensure that the functional performance of ME EQUIPMENT meets established criteria and to enable the early recognition of changes in the properties of components of the ME EQUIPMENT, and to verify compliance with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning.
This document also contains requirements for the ACCOMPANYING DOCUMENTS associated with ACCEPTANCE AND CONSTANCY TESTING of the DENTAL CBCT EQUIPMENT.
This document does not apply to:
- aspects of thermal, EMD (electromagnetic disturbances), mechanical and electrical safety;
- aspects of mechanical, electrical and software performance, unless they are essential for performing the ACCEPTANCE TESTS and CONSTANCY TESTS, and directly affect image quality, RADIATION OUTPUT and PATIENT positioning.
Equipment in the scope of IEC 61223-3-5 is excluded from the scope of this document.
DENTAL EXTRA-ORAL X-RAY EQUIPMENT can provide modalities which are in the scope of IEC 61223-3-4. In this case, the respective clauses of the IEC 61223-3-4 apply.
The object of this document is to establish:
- the essential parameters which describe the performance of DENTAL CBCT EQUIPMENT with regard to the image quality, RADIATION OUTPUT and PATIENT positioning;
- methods of testing and whether measured quantities related to those parameters comply with the specified requirements.
These methods rely on non-invasive measurements performed once the installation or a MAJOR SERVICE ACTION is completed.

Best Sellers From Health Sector

GSO 2017:2023
 
Gulf Standard
Personal Care and Baby Wet Wipes
BH GSO 2017:2023
GSO 2017:2023 
Bahraini Standard
Personal Care and Baby Wet Wipes
BH GSO 2241:2023
GSO 2241:2023 
Bahraini Standard
Cosmetic Products - Body Cleaners
GSO 2241:2023
 
Gulf Standard
Cosmetic Products - Body Cleaners

Recently Published from Health Sector

GSO ISO 12219-9:2025
ISO 12219-9:2019 
Gulf Standard
Interior air of road vehicles — Part 9: Determination of the emissions of volatile organic compounds from vehicle interior parts — Large bag method
GSO ASTM E3032:2025
ASTM E3032:22 
Gulf Standard
Standard Guide for Climate Resiliency Planning and Strategy
GSO ASTM E3356:2025
ASTM E3356:22 
Gulf Standard
Standard Guide for Stakeholder Engagement on Environmental Risk Management and Climate
GSO ISO 12219-1:2025
ISO 12219-1:2021 
Gulf Standard
Interior air of road vehicles — Part 1: Whole vehicle test chamber — Specification and method for the determination of volatile organic compounds in cabin interiors